Home Precision Environment Particle Measurement CPA 341 Cleanroom Particle Analyzer
CPA 341 Cleanroom Particle Analyzer product image thumbnail
Topas GmbH Cleanroom Particle Counter

CPA 341 Cleanroom Particle Analyzer

Portable or stationary single-particle light-scattering counter for low-concentration aerosols in gaseous samples — covering 0.3 to 10 µm at a fixed sampling flow of 28.3 L/min for cleanroom qualification, acceptance testing and filter measurements.

Size range
0.3 – 10 µm
Size classes
5 fixed particle-size classes
Sample flow
28.3 L/min (1 cfm)
Operation
Touchscreen with up to 999 programs
28.3 L/min 0.3–10 µm 5 size classes Mobile / stationary Touchscreen
ISO 14644-1 ISO 14644-3 GMP Annex 1 VDI 2083 ISO 21501-4

Model Overview

Model Measurement principle Core specification Core fit
CPA 341 Single-particle 90° light-scattering counter with internal flow generation 0.3–10 µm, five fixed size classes and 28.3 L/min sample flow Cleanroom qualification, acceptance measurements, filter-integrity support and fractional-efficiency testing

Key Features

Stable 28.3 L/min sampling

An internal side-channel blower maintains a 28.3 L/min (1 cfm) sampling flow, supporting both mobile measurements and stationary monitoring of low-concentration aerosols.

Five fixed particle-size classes

The optical counter measures particles from 0.3 to 10 µm in five preset size classes, providing particle-number and size information for cleanroom and filter-test workflows.

Programmable cleanroom workflows

Up to 999 measurement programs can store measurement and pause times, measurement positions, cycle counts and cleanroom-class alarm settings for repeatable qualification routines.

Touchscreen, alarms and data handling

The glove-operable touchscreen supports single, repeat and continuous measurement modes. Results can be displayed and printed, transferred through Ethernet and processed with CRQWin software.

Technical Specifications

Parameter Unit Value / Description
MEASUREMENT SYSTEM
Measurement principle Single-particle light scattering with 90° collection optics
Particle-size range µm 0.3 – 10
Number of size classes 5, fixed
Sample flow L/min 28.3 (1 cfm)
Flow generation Internal side-channel blower
Zero-count rate counts < 1 count per 5 minutes, evaluated according to ISO 21501-4
Counting efficiency % 50% ± 20% at 0.3 µm, evaluated according to ISO 21501-4
OPERATION AND DATA
Display and control Touchscreen
Stored measurement programs Up to 999
Measurement modes Single, repeat and continuous measurement
Measurement / pause time 1 second to 99 hours
Alarm Configurable for each cleanroom class
Interface Ethernet
Result handling Tabular display, thermal-printer output and CRQWin processing
Inlet connection mm Ø 8
PHYSICAL AND ELECTRICAL
Power supply 100 – 240 V AC, 50/60 Hz
Dimensions (W × H × D) mm 225 × 255 × 220
Weight kg 4

Specifications are based on the manufacturer data sheet dated March 2026. Confirm the final delivery scope, accessories and calibration documentation before ordering.

Applications

Classification and validation

Particle-concentration measurements for cleanroom qualification and validation according to ISO 14644-1/-3, GMP Annex 1 and VDI 2083.

Cleanrooms and workbenches

Mobile acceptance measurements of cleanrooms, clean-air workbenches and ventilation systems with high cleanliness requirements.

Integrity and fractional efficiency

Upstream concentration measurement and downstream leak scanning in filter-integrity workflows, plus determination of fractional separation efficiency for filters and separators.

Related Particle Measurement Products